ASTM D1298

View as Grid List

1 Item

Set Descending Direction
  1. Tamson D1298 Manual density
    ASTM D1298 density apparatus
per page

Test

The sample is brought to the prescribed temperature and transferred to a cylinder at approximately the same temperature. The appropriate hydrometer is lowered into the sample and allowed to settle. After temperature equilibrium has been reached, the hydrometer scale is read, and the temperature of the sample is noted. If necessary the cylinder and its contents nnay be placed in a constant temperature bath to avoid excessive temperature variation during the test.

Precision

Product

Temperature range

Units

Repeatability

Reproducability

Transparent

-2to24.5°C

Density

0.0005

0.0012

Nonviscous

29 to 76°F

Specific gravity

0.005

0.0012

 

42 to 78°F

API Grafity

0.1

0.3

Opaque

-2to24.5°C

Density

0.006

0.0015

 

29to76°F

Specific Gravity

0.0006

0.0015

 

42to78°F

API Gravity

0.2

0.5